Akta Pilot Ph Probe

Sanitization of ÄKTA pilot 600 | Cytiva, ÄKTA pure | Cytiva, ÄKTA ready single-use system | Cytiva, ÄKTA pure | Cytiva, The pH probe should be calibrated prior to each use. If outside of the specification, >80% slope and +/-60 mV asymmetry, replace the probe . Check the pH electrode for cracks or damage and consider replacing it once every six months. Change the pump rinsing solution frequently. The rinsing solution should be 20% ethanol in deionized water.

ÄKTA pilot 600 Fig 1. ÄKTA pilot 600 system is available in two versions, 600S for non-GMP and 600R for GMP environments. Thanks to the modular design, both versions can easily be modified to match specific needs at different times. Fig 2. ÄKTA pilot 600 helps simplify operations from column packing to scale up and technology transfer.

ÄKTA pure chromatography system allows for quick and easy customization thanks to the modular design of both the hardware and software. Choose from ÄKTA pure 25 with a flow rate up to 25 mL/min or ÄKTA pure 150 with a flow rate up to 150 mL/min and combine with more than 20 hardware options. Control gradients and elution conditions with pH …